CArmSystem
SPECIFICATIONSOFLAPAROSCOPICSET
Group-A(ImagingandInsufflationdevices)
Sno.
Detailsoftheinstruments
Nos.
1.
3ChipCameraSystem,ColorSystem,PALcomprisingof(a)and(b)
(a)CameraControlunitwithintegrateddigitalimageprocessingmodule.Itmustbesuppliedwithallstandardaccessorieslikemainscord,keyboard,connectingcableoflength.175cms,BNC/BNCvideocableSvideo(Y/C)connectingcables,RGBconnectingcable,DVcable.Itmusthavefollowingfeatures
∙Videooutputs–CompositesignalatBNCsocket,S-VideoSignalto4pol.MiniDin.Socket(2x),RGBSignalto4xBNCsockets,DVSignalto6pinDVsocket
∙Inputs-Keyboardinputfortitlegeneratorandcamerafunctionsto5pinDINsocket
∙PowerSupply-100-240VAC,50/60Hz
∙Certifiedto–IEC601-1,601-2-18,CSA22.2No.601L2601andCEaccordingtoMDDprotectionclass1/BF
1
(ONE)
(b)Threechipcameraheadofcolorsystem,integrated,ParfocalZoomlens,f25-50mm(2x),with2freelyprogrammablecameraheadbuttonsincludingplasticcontainerforsterilizationandstorage.Itmusthavefollowingfeatures:
-
∙ImageSensor–3x½inchCCDchip
∙Pixels–752(H)X582(V)pixelsperchip(PAL)
∙Resolution–min.750lines(Horizontal)
∙SignaltoNoiseRatio->60dB
∙AGC–Microprocessorcontrolled
∙Min.Sensitivity–ParfocalZoom,Cmount3Lux(f=1.4mm)
∙Exposurecontrol–1/50secto1/10000sec(PAL)
1
(ONE)
2.
19“flatscreenmonitor,Desktopversion,Colorsystem,PAL,resolutionmax.1280X1024SDI,Composite,S-Video,RGB,DVIandS-XGAInputBrightness,450cd/m2,Contrast650:
1PowerSupply,100-240VAC,50/60Hz.
1
(ONE)
3
a.ColdLightFountainXENONwithone300watt,15VXenonlampPowersupply220-240VAC,withcolortemperatureupto6000Kwithcontinuousmanualadjustmentoflightintensityincludingmainscord.
b.SpareXenonebulb300watt
1
(ONE)
1(One)
4
ElectronicCO2insufflatorwithGasflowof19litrs/minuteormorewithtouchkeysfordisplayofatleast4parameters(Insufflation,pressure,intra-abdominalpressure,gasflowandgasconsumption)settingswithopticalandacousticalarmsforpatientoverpressure.Itshouldhaverefillingsafetysystemwithreleaseofpressureincaseofoverpressureandwithautomaticelectronicallycontrolledgas.Tobesuppliedwithmaincord,siliconsterilizabletubingsetsUniversalwrench.
1
(ONE)
5.
LightGuideCable–4.8to5.0mmdia,>200cmlength,autoclavable,compatablewithlightsource&telescope.
2
(TWO)
6.
Telescope:
a.Telescope30cminlength,autoclavable,wideangleview,10mmdia,0degreeforwardviewing
b.Telescope30cminlength,autoclavable,wideangleview,10mmdia,30degreeforward–obliqueviewing
c.Telescope30cminlength,autoclavable,wideangleview,5mmdia,30degreeforward–obliqueviewing
2
(TWO)
1
(ONE)
1
(ONE)
Group-B(LaparoscopicInstrumets)
Sno.
Detailsoftheinstruments
Nos.
1.
TrocarsandCannulae-
a.Reusable,pyramidicaltiptrocarsandcannulaautomaticandmanuallyoperated,exchangeablefunctionaldoorvalveandguidetube,maintenancefreestopcockforgasinsufflationofsize11mm
b.Reusablepyramidicaltiptrocarsandcannulaautomaticandmanuallyoperated,exchangeablefunctionaldoorvalveandguidetube,maintenancefreestopcockforgasinsufflationofsize6mm
c.Endotipcannula11mmoflangth>10cmwithmultifunctionalvalve.
5
(FIVE)
6
(SIX)
1
(ONE)
2.
VeressNeedle:
a.Veressneedlespringloadedof130mmlength
b.Veressneedlespringloadedof150mmlength
2(TWO)
1(ONE)
3.
a.Reducersleeveforconversionof10/11mmto5/6mm
b.Reducer11/5mm
2(TWO)
2(TWO)
4.
Laparoscopichandinstruments–
a.DissectingandGraspingForcepsautoclavablerotatable,dismantalable,insulatedinstrumentswithH-Fconnectorspreferablywithdoubleactionjawsof5mmdiameterand36cmlengthwithergonomichandle(withorwithoutrachetasperfunction).
1.Kelley’sdissectingandgraspingforceps
2.Kelley’sdissectingandgraspingforcepswithlongjaws
3.RightangledKelley’sdissectingandgraspingforceps
4.Graspingforceps,atraumaticfenestratedandcurvedwithlongjaws
5.Graspingforcepsatraumaticwithlongjaws
6Graspingforcepswithfenestratedjaws
7.Graspingforcepswithmultipleteethwithjawsof4.8mmwidth
8.Graspingforcepswithtigerjaws2X4teeth
9.Babcocksgraspingforcepsatraumaticfenestrated
10.Babcocksgraspingforcepsatraumaticjawswithmultipleteethfenestratedlong
11.Clawforceps2X3teeth
l.2Bowelgrasperfenestrated
13.Forcepsforadenolysis
2(TWO)
1(ONE)
1(ONE)
2(TWO)
2(TWO)
2(TWO
2(TWO)
2(TWO)
2(TWO)
1(ONE)
1(ONE)
2(TWO)
1(ONE)
1(ONE)
b.Scissors:
Autoclavable,rotatable,dismantalable,insulatedinstrumentswithH-Fconnectorspreferablywithdoubleactionjawsof5mmdiameterand36cmlengthwithergonomichandle.
1.MetzenbaunScissors12mmblade
2.MetzenbaunScissors17mmblade
3.Blakesleysdissectingandbiopsyforceps
2(TWO)
1(ONE)
1(ONE)
c.Instrumentsforcuttingandcoagulation(Unipolar/Bipolar)
1.Dissectingspatula5mm
2.2DissectingHookcannulaLshaped
3.3HighfrequencyUnipolarcord
4.4Bipolarcoagulationdismantalableforcepswithfacilityofchanging,insert
5.5Bipolargraspingdismantalableforcepswithfacilityofchanging,insert
6.6AssortedInsertsforbipolarforceps
7.7Highfrequencybipolarcord
2(TWO)
3(THREE)
5(FIVE)
1(ONE)
1(ONE)
4(FOUR)
4(FOUR)
Group-C(ClipApplicators,NeedleHolderandSuturingDevices)
d.Needleholdersandsuturingdevices:
1.Needleholderergonomichandlewithratchetleftcurvedjawswithtungustencarbideinserts5mmdiameter&length33cm
2.AssistantNeedleholderFlamingoJawstraighthandlewithratchetsize5mmlength33cm
3.Fascialclosureinstrument
4.Integralknotpusherandthreadcatcher
5.GBspeculam
1(ONE)
1(ONE)
1(ONE)
1(ONE)
1(ONE)
e.Clipapplicator
1.Clipapplicatorforusewithethicontitaniumclips(mediumlarge)dismantlingrotatingwithratchettolockthejawpartholdingtheclipsize10mmlength36cm
2.Clipapplicatorforusewithethicontitaniumclips(medium)dismantlingrotatingwithratchettolockthejawpartholdingtheclipsize10mmlength36cm.
1ONE)
1(ONE)
Group-D(SuctionandIrrigationdevices)
f.Suctionandirrigationdevices
1Cannulaforpunctureandaspiration
2Suction/irrigationtubewithmultileholesandtwowaystopcock5mm
3Suction/irrigationtubewithmultileholesandtwowaystopcock10mm
4Suctionirrigationpump–suctionandirrigationpump,highflowsystemwith3L/mincapacitywithalltubingforsuctionandirrigationcannulaandjarforfluidofsuction
2(TWO)
2(TWO)
1(ONE)
1(ONE)
Group-E(Miscellaneous/Others)
h.Miscellaneous
1CO2cylinderwithhighpressurehose–bulktype
2CO2cylinderwithhighpressurehose–3KGmedicalgradecylinder
3CVTof1KVA
4.Trolley–mobiletrolleyforkeepinglaparoscopicequipmentswithsufficientnumberofpowersockets
1(ONE)
1(ONE)
1(ONE)
1(ONE)
Note:
Tenderersmayquoteanyorallgroups.Ernestmoneyofgroupstenderedmustaccompanywithtenders.
CHOLEDOCHOSCOPESET
Itmustcompriseoffollowing:
(A)CHOLEDOCHOSCOPE(2mmchannel):
One
-Fullyimmerssibleindisinfectantsolution.
-Accurateobservationbackedupbyincreasedimagesize.
-Increasedimageguidefiberdensity.
-Widesuctionchannel
FieldofView:
120degreeormore(forwardviewing)
Workinglength:
375mmormore
DistalEnddia.:
5.0mmorless
Insertiontubedia:
4.9mmorless
DepthofField:
3mmto50mmorbetter
BendingRange:
Up–160degree
Dn–130degree
Channeldia.:
2.2mmormore
Biopsyforcepsminimalvisibledistance:
5mmorlessfromdistalend.
(B)SINGLECHIPCCTVCAMERA:
One
10bitdigitalsignalprocessingwith¼inchCCDchip.
InbuiltMoirefilterforcompatibilitywithfiberopticendoscopes.
Resolution:
470horizontallines.
Signaltonoiseration:
Morethan50db
Detachableandlockableadaptorwithfocusingfacility.
Videooutput:
Y/c&Composite.
Fullyimmerssiblecamerahead&cableassembly.
(C)COLOURMONITOR:
One
14inchmedicalgradeLCDMonitor.
(D)HALOGENLIGHTSOUCE:
One
-Itshouldbecompactandlightweightaround5–6kgorlessforeasiertransportability.
-Shouldhaveautoclavablewaterbottle.
-Shouldhave150wattshalogenlampwithstandbylampoption.
-ShouldhaveBuiltinAir-pumpfordistention&automaticleakagetesting.
-Shouldbecompatiblewithrigidandflexibleendoscopesboth,
-Pumppressure0.3–0.6kg/cm2(atOcc./min.)0.18kg/cm2orbelow(at2000cc/min.)
(E)Accessories
(a)LeakagetesterOne
(b)CompatibledormiawirebasketOne
(c)CompatiblebiopsyforcepsTwo
(d)Choledeoscopegraspingforceps5mmOne
(foruseinLaparoscopysurgery)
SpecificationforRadioFrequencyAblationSystem
TheSystem
oShouldhaveRadioFrequencygeneratorwithfeedbackalgorithmforcontinuoustissueimpedancemonitoringandautomaticoutputadjustment.
oShouldhaveautomaticcurrentandimpedancecontrolfeature.
oShouldhave200-wattpowersourcewithpulsatingenergyforgeneratinglargercoagulationvolume.
oShouldhavethermocoupletomeasurecoagulationzonetemperatureandasthetemperaturecycleiscompleteelectrodemayberepositionedtomeasurecoagulationzonetemperature.
oShouldhaveautomatedtreatmentcycletohavesuccessfultypicaltissueablationinmaximum12minutes.
oShouldhaveLCDstatuswindowtohaveuser-friendlyfeedbackduringsetup